Customization: | Available |
---|---|
Certification: | Ce, RoHS, FCC, ISO, SGS Certificates |
Shock Range: | -60c~+150c |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model | ZH-TS-42 (A~C) |
ZH-TS-60 (A~C) |
ZH-TS-80 (A~C) |
ZH-TS-100 (A~C) |
ZH-TS-120 (A~C) |
ZH-TS-150 (A~C) |
Inner size: WHD (cm) | 40*30*35 | 50*30*40 | 50*40*40 | 50*50*40 | 60*40*50 | 60*50*50 |
External size: WHD (cm) | 150*180*150 | 160*175*160 | 160*185*160 | 160*185*170 | 170*185*170 | 170*195*170 |
Temperature range (Test zone) | High temperature:+60ºC~+200ºC; low temperature: -10ºC~-65ºC( A:-45ºC; B:-55ºC; C:-65ºC) | |||||
Heating Time | RT~200ºC about 30min | |||||
Cooling Time | RT~-70ºC about 85min | |||||
Temp. Conversion Time | Less than 10S | |||||
Temp. Recovery Time | Less than 5min | |||||
Temp. Deviation | ±2.0ºC | |||||
Temp. Fluctuation | ±0.5ºC | |||||
Material | External material: SUS#304 Stainless steel plate Internal material: SUS#304 Stainless steel plate |
|||||
Set Mode | Chinese / English menu, touch button input | |||||
Controller | Tato in Taiwan | |||||
Cooling System | Water-cooled or air-cooled, Imported GEA Compressor in France | |||||
Protection devices | Fuse switch, compressor overload switch, refrigerant high and low pressure, protection switch, super humidity over tempe. protection switch, fuse, failure warning system | |||||
Patrs | Viewing window; 1 Φ50mm (Φ100mm) access port or more available; 2 layers or more available 4 flexible footwheels |
|||||
Power | AC380V 50/60Hz Three-phase four-wire, AC power | |||||
weight (kg) | 750 | 790 | 830 | 880 | 950 | 1050 |